Etd

A Minimum-Bending-Energy Needle Model for Closed-Loop Localization During Image-Guided Insertion

Pubblico

Contenuto scaricabile

open in viewer

Accurate needle placement is critical to the success of needle-based interventions. Needle deflection due to tissue non-homogeneity and dynamic forces results in targeting error, potentially requiring repeated insertions. Real-time imaging enables closed-loop control of the needle during insertion, improving insertion accuracy. The needle localization algorithm proposed in this thesis models the needle as a parametric polynomial equation optimized to minimize beam bending energy relative to a set of observed needle coordinates. Simulated insertions using an MRI dataset show that the minimum bending energy model allows planning of subsequent imaging planes to capture the moving needle while estimating the shape of the needle with low error.

Creator
Contributori
Degree
Unit
Publisher
Language
  • English
Identifier
  • etd-042518-145218
Parola chiave
Advisor
Committee
Defense date
Year
  • 2018
Date created
  • 2018-04-25
Resource type
Rights statement

Relazioni

In Collection:

Articoli

Elementi

Permanent link to this page: https://digital.wpi.edu/show/z890rt41f