Reflectance-Based Pulse Oximetry for the Chest and Wrist
PublicDownloadable Content
open in viewerReflectance-based pulse oximetry is a technique used for noninvasively monitoring the oxygen saturation (SpO2) and pulse rate (PR). However, there is little supporting evidence that it can accurately collect measurements from the chest and wrist. In this project, a reflectance-based pulse oximeter was built and used to collect measurements while sitting, standing, during self-induced hypoxia, and during self-induced hyperventilation then compared to the measurements taken by a HOMEDIC Deluxe Pulse Oximeter. The prototype was able to accurately measure within an error of + 1% and ±3% for SpO2 and PR respectively from the wrist while an error of ±1% and +4% for SpO2 and PR respectively from the chest.
- This report represents the work of one or more WPI undergraduate students submitted to the faculty as evidence of completion of a degree requirement. WPI routinely publishes these reports on its website without editorial or peer review.
- Creator
- Publisher
- Identifier
- E-project-042413-160152
- Advisor
- Year
- 2013
- Date created
- 2013-04-24
- Resource type
- Major
- Rights statement
Relations
- In Collection:
Items