Student Work

Sound Pulse TM

Public

Downloadable Content

open in viewer

Strokes are a major cause of death and disability in the US, with ischemic strokes being the most common. This Major Qualifying Project's (MQP) goal was to create a device that can detect disrupted blood flow in the carotid artery, which is an indication of plaque build-up that can lead to ischemic strokes. Based on the previous team’s research, our team developed a pump, neck model, and arteries in a closed-loop system to test various sensors. Through testing on the model, we found a suitable method for detecting plaque buildup. The team’s most promising results came from testing with piezoelectric sensors encased in silicone housing.

  • This report represents the work of one or more WPI undergraduate students submitted to the faculty as evidence of completion of a degree requirement. WPI routinely publishes these reports on its website without editorial or peer review.
Creator
Subject
Publisher
Identifier
  • 64531
  • E-project-042722-163414
Keyword
Advisor
Year
  • 2022
Sponsor
UN Sustainable Development Goals
Date created
  • 2022-04-27
Resource type
Major
Rights statement

Relations

In Collection:

Items

Items

Permanent link to this page: https://digital.wpi.edu/show/t722hd13x